Draw a box in an even area of the parent metal or weld. Clearshot measures the mean and noise and normalises with your detector's basic spatial resolution (SRb), giving SNRN. It compares this with the minimum in ISO 17636-2:2022 Table 3 (steel, copper and nickel alloys) or Table 4 (aluminium and titanium) for your source, energy, thickness and testing class.
| Source | Class A | Class B |
|---|---|---|
| X-ray U ≤ 50 kV | 100 | 150 |
| X-ray 50 kV < U ≤ 150 kV | 70 | 120 |
| X-ray 150 kV < U ≤ 250 kV | 70 | 100 |
| X-ray 250 kV < U ≤ 1 000 kV | 70 | 100 |
Higher energies and gamma sources have thickness bands; Clearshot holds every row of both tables, the HAZ measurement factor and the clause 6.9 exception for gamma sources. Other agreed values, such as the 7.3.1 exception, go in as a manual minimum with the reason recorded.